Solutions / Tympanometers / OAE Testing with PE Tubes
OAE Testing with PE Tubes
OAE Testing with PE Tubes
Learn how to perform OAE testing with PE-tubes using the Titan.
Titan Support
This quick guide describes the procedure when wishing to perform an OAE test on a patient with PE-tubes.
When performing an OAE test the in-ear detection ensures that that the probe is properly inserted in the ear, as well as calculates the correct stimuli level for the testing.I the patient has PE-tubes inserted, the Titan will in many cases indicate out-of-ear, because of the hole in the eardrum. In this case, follow the procedure as described below.
Turn on Titan by pressing the R or L button.
Select the desired OAE test protocol from the dropdown list.
Select the ear for testing.
The ideal test environment is a quiet room. High ambient background noise adversely affects OAE measurements.
The patient should be sleeping or in quiet relaxed state. Talking, swallowing, blinking, crying or movement may affect testing.
Place the probe in the ear – the light will turn green when a good seal is obtained.