Welcome to the product training for AC40 audiometer. On this page you will find an overview of all the available training materials and support.
The short increment sensitivity index (SISI) test measures the ability to recognise 1dB increases in intensity during a series of bursts of pure tones presented 20dB above the pure tone threshold for the test frequency. It can be used to differentiate between cochlear and retro cochlear disorders, as a patient with a cochlear disorder will be able to perceive the increments of 1dB, whereas a patient with a retro cochlear disorder will not.
Headphones or insert phones
Press and hold the Tests button and select SISI: Short increment sensitivity index using the scroll wheel on the audiometer.
Select Tone or Warble in channel 1. If cross hearing is suspected, masking should be presented in channel 2.
Select the desired test frequency and set the input level 20dB above threshold using the scroll wheel on the audiometer.
In the most common type of SISI test, the incremental steps are set to 1dB. By pressing and holding the function button AM, it is possible to change the increments. Prior to presenting the actual test of 1dB increments, a trial of 2 or 5dB steps can be employed to ensure that the patient understands the test procedure and what will be required. If you are uncertain about the responses from the patient, 0dB can be used.
Explain to the patient that they will now hear a series of tones. If suddenly there is a change in the loudness of the presentation of the tone, the response button should be pushed.